National standard spectrophotometric water quality online monitoring instrument

Residual chlorine (total chlorine) online monitor

Online nitrate monitor

Online phosphoric acid water quality monitoring instrument

Online silica monitor

Online fluoride monitoring instrument

Hydrazine online monitor

CODcr Water Quality Online Monitor

Ammonia nitrogen water quality online monitor

Online monitoring instrument for total nitrogen water quality

Online monitoring instrument for total phosphorus water quality
Service Hotline
+86-0632-2723255 +8618363281129
A mixture of water sample, potassium dichromate digestion solution, silver sulfate solution (silver sulfate as a catalyst can more effectively oxidize straight-chain fatty compounds), and concentrated sulfuric acid is heated to 175°C. After oxidizing the dichromate ions, the color of the organic matter in the solution changes. The analyzer detects this color change and converts this change into a COD value output. The amount of dichromate ions consumed is equivalent to the amount of organic matter oxidizable by oxidation, i.e., COD.
Technical Specifications |
|
Measurement methods | Potassium dichromate method |
Implement standards | GB/T11914-89 ; HJ 377-2019 |
Measurement range | 15.0~5000.0 mg/L (range customizable) |
Measurement cycle | 50 minutes |
Precision | ≤ 5% F.S. Use a 50% range standard solution |
Application industries | Industrial water process monitoring in the power, chemical, and steel industries, as well as water quality monitoring of wastewater, drinking water, and surface water. |
Automatic cleaning | Yes |
Ambient temperature | 5~28℃ |
Relative humidity | 5~95% |
Sample requirements | Turbidity <50 NTU, maximum particulate size 100 μm< 0.1 g/L |
Sampling method | Through an external overflow container |
Reagent consumption | 260 or 500 times |
Power supply | 110-240 VAC、4A,50/60 Hz |
Maximum power consumption | 150 VA |
Analog signal output | 4-20 mA, maximum load of 500Ω |
Digital signal output | Supports RS485 (Modbus RTU) / Ethernet (Modbus TCP) |
Protective enclosure | IP67 |
Dimensions (H*D*W) | 690*370*400mm |
Weight | 30kg |
Copyright © Mozhidao (shandong) Measurement & Control Equipment Co,Ltd
Address:No. 267, 2nd Floor, East District, Zhimao City, Honghe Road, Jinghe Street, Tengzhou City, Zaozhuang City, Shandong Province
Email:mzd18363281129@163.com
Technical Support:tengzhou.com.cn
Service Hotline
WeChat Official Account
Tiktok