National standard titration method water quality online monitoring instrument

Online monitoring device for water alkalinity (methyl orange alkalinity).

Online monitoring device for water quality alkalinity (full alkalinity).

Online chloride ion monitor

Online water hardness monitoring instrument

Online monitoring instrument for water hardness (high concentration).

Online calcium hard monitoring device

Online sulfate monitoring instrument

Online Methyl Orange Acidity Monitor

Online Phenolphthalein Acidity Monitoring Instrument

Permanganate index water quality online monitor
Service Hotline
+86-0632-2723255 +8618363281129
The principle of the titration method: The online alkalinity monitoring instrument is based on the acid-base titration method. This instrument uses a precise sampling system to introduce the water sample to be tested into the reaction unit. The water sample reacts with a specific acid standard solution. As the titrant is added drop by drop, the alkaline components in the sample are gradually neutralized, and the color of the reaction solution also changes. When the color change reaches the preset endpoint, it is considered that the alkaline substances in the sample have been completely consumed. At this point, the instrument calculates the alkalinity value of the water sample based on the titrant consumption and reaction conditions.
Technical Specifications |
|
Measurement methods | National standard titration method |
Implement standards | GB/T14419-2018 |
Measurement range | 0.00~5.00 mmol/L (phenolphthalein basicity) |
Measurement cycle | ≤ 15 minutes |
Precision | ≤ 5% F.S. Use a 50% range standard solution |
Application industries | Industrial water process monitoring in the power, chemical, and steel industries, as well as water quality monitoring of wastewater, drinking water, and surface water. |
Automatic cleaning | Yes |
Ambient temperature | 5~28℃ |
Relative humidity | 5~95% |
Sample requirements | Turbidity <50 NTU, maximum particulate size 100 μm< 0.1 g/L |
Sampling method | Through an external overflow container |
Reagent consumption | 260 or 500 times |
Power supply | 110-240 VAC、4A,50/60 Hz |
Maximum power consumption | 150 VA |
Analog signal output | 4-20 mA, maximum load of 500Ω |
Digital signal output | Supports RS485 (Modbus RTU) / Ethernet (Modbus TCP) |
Protective enclosure | IP67 |
Dimensions (H*D*W) | 600*300*500mm |
Weight | 30kg |